name: mri-hardware
description: >-
  Read before any MR hardware or safety calculation: B0 field maps and
  shimming with current limits, gradient and RF coil fields, B1+rms and SAR
  (IEC 60601-2-33), dB/dt and nerve stimulation, receiver noise and SNR, low-
  field and open-source consoles. Prevents silent errors in units, frames,
  field components, harmonic normalization, bounded optimization and limit
  definitions. Ships tested SI helpers checked against analytic cases. Not
  clinical advice.
metadata:
  author: Ke Wang
  version: "0.13.0"
---

# MRFoundry: MRI Hardware and Safety

Hardware numbers are useful only with their units, conventions and a check. Use the tested helpers in
`scripts/mrfoundry_hwlib.py` (numpy/scipy, SI units) and import them rather than re-deriving them:
`python .agents/skills/mri-hardware/scripts/mrfoundry_hwlib.py --api` lists signatures (do not cat the source); `scripts/test_hwlib.py` shows
each helper on an analytic case.

## Route

| Task | Use (`mrfoundry_hwlib.py`) |
|---|---|
| Field map from multi-echo phase (nonzero phase0, uneven echo times) | `fieldmap_multiecho(phase, te, magnitude, max_abs_hz)` (temporal unwrapping; also spatially unwrap if fields can exceed 1/(2 ΔTE_min)) |
| Shim basis at voxel positions | `voxel_world_coords(affine, shape)` (full affine, mm to m), then evaluate the vendor's polynomials there; `solid_harmonics`, `fit_harmonics` for analysis |
| Optimal shim currents within limits | `shim_currents(field, roi, channel_maps, lower, upper, current_during_acquisition=I_acq)`: absolute currents, `frequency_adjust` (= -offset) and a KKT optimality report |
| Any bounded least-squares fit you wrote yourself | `bounded_lsq_kkt(A, b, x, lower, upper)` must return ok |
| Coil fields, gradient efficiency and linearity | `biot_savart`, `circle_path`, `loop_bz_on_axis`, `gradient_metrics` |
| RF exposure | `rf_energy`, `b1_rms` (window includes dead time), `b1_rms_of_train`, `b1_for_flip`, `sar_scale`, `iec_sar_limits`, `time_averaged_sar` |
| Gradient safety | `slew_rate`, `dbdt_at_radius`, `effective_stimulus_duration`, `pns_threshold_dbdt`, `iec_dbdt_limit` |
| Safety numbers of an exported Pulseq file (B1+rms, B1 peak, gradient rms, ramp PNS, crusher moments) | `pulse-sequence-design/scripts/check_exported_seq.py seq --safety --pns-rheobase .. --pns-chronaxie ..` (limits: `--b1rms-max`, `--pns-max`, ...) |
| Off-resonance correction of a long readout with a field map | `fieldmap_multiecho` here, then `mri-reconstruction` `offres_least_squares` and `smooth_field` |
| Receiver noise and SNR | `noise_covariance`, `prewhitening_matrix`, `roemer_combine` (SNR units), `rss_combine`, `snr_scale` |

Derivations, standard values and references: [computations](references/computations.md).

## Checks that prevent silent errors

1. **Write the convention down before computing**: units (T, mT, µT; Hz, ppm; mT/m vs G/cm; per ampere),
   frame and axis (z = bore axis; world RAS+ from the NIfTI affine), field component (Bz vs |B|), region of
   interest, and which limit (6-min vs 10-s SAR, peak vs RMS, normal vs first-level mode).
2. **Field-model bookkeeping for shimming**: the measured map already contains the field of the currents that
   were on during the acquisition. Solve for ABSOLUTE new currents with `current_during_acquisition`, bound
   those, and map the free offset to the centre-frequency convention you report.
3. **Prove optimality, do not assume it**: run `bounded_lsq_kkt` on any constrained solution (column scaling
   left unscaled, bounds applied to scaled variables and sign flips all pass a "residual went down" check).
   Report which channels sit at their limits.
4. **Validate the tool on an analytic case first** (loop on axis, solenoid, synthetic harmonic field, a
   rectangular RF train) and check discretization convergence before the real geometry.
5. **Use established solvers for geometry-dependent physics** (Biot-Savart for coils; full-wave EM with a
   body model for local SAR; published PNS models). Never present a uniform-body estimate as local SAR.
6. **Design inside limits with a margin** (2-5 %): size flips, pulse durations, echo counts and TR from the limit
   before building (`b1_rms_of_train`), because rounding to rasters and ramp shapes move realized values; recompute
   every reported safety number from the exported file. Simulated signals of an exported sequence come from KomaMRI
   with the replay check of `pulse-sequence-design`.
7. **Report margins and limits**: standard and edition, mode, averaging window, assumptions and what was not
   modelled. Say when a qualified MR safety expert is required.

## Before you finish

    python .agents/skills/mri-research-workflow/scripts/finish_check.py --require <every deliverable name> \
        --rerun "python <entrypoint> <documented arguments with {inputs} and {out}>" --inputs <input folder>

## More (open only when needed)

- Hardware chain, low-field and open-source consoles, coil and shim design tools, MR safety references:
  [hardware landscape](references/landscape.md) and the hub's [hardware guide](../mri-research/references/hardware.md).
- Citations: [reading list](references/reading-list.md). Setup problems:
  [tool setup guide](../mri-research/references/tool-setup.md). Project notes (only with `.mrfoundry/INDEX.md`):
  [project memory](../mri-research/references/project-memory.md).
- Waveforms: `pulse-sequence-design`. Reconstruction: `mri-reconstruction`.